Abstract

좌심실보조기의 임상응용범위를 넓히기 위해서 개심술을 하지 않고 좌심실보조기를 적용할 수 있도록 캐뉼라형 좌심실보조기를 이용한 보조순환 실험을 하였다. 좌심실 보조기는 모터의 회전력이 PVC 벨로우즈를 수측, 팽창시키는 전기유압식으로 구동되며 환자는 양쪽 대퇴동맥을 통하여 보조순환을 받도록 되어 있다. 두차례의 동물실험을 통하여 대퇴동맥을 통한 보조 순환방식의 가능성 및 그 효과에 대해서 관찰하였다. 첫 실험은 12kg 개를 이용하여 셀딩거 방법으로 뮬린형 카테터를 디퇴동맥으로 삽입히여, 캐뉼라형 보조순환방법의 가능성을 확인하였고, 두번째 동물실험에서는 내경 4.8 mm의 PVC 카테터를 이용하여 좌심방에 유입카테터를 연결하고 대동맥에 유출카테터를 연결하여 보조순환에 의한 대동맥압의 변화와 혈류량의 변화를 관찰하였다. 보조순환동안 개의 동맥압을 100mmHg로 유지할 수 있었으며, 좌심실내압이 감소하는 것을 관찰할 수 있었다.