임피던스 혈량측정법을 이용한 손가락 혈류량의 측정

Development of Impedance Plethysmography and Measurement of Digital blood Flow

  • 김덕원 (연세대학교 의과대학 의용공학과) ;
  • 송철규 (연세대학교 의과대학 의용공학과) ;
  • 김원기 (연세대학교 의과대학 의용공학과) ;
  • 이명호 (연세대학교 공과대학 전기공학과)
  • Kim, Deok-Won (Dept. of Medical Eng., College of Medicine, Yonsei Univ.) ;
  • Song, Chul-Gyu (Dept. of Medical Eng., College of Medicine, Yonsei Univ.) ;
  • Kim, Won-Ky (Dept. of Medical Eng., College of Medicine, Yonsei Univ.) ;
  • Lee, Myong-Ho (Dept. of Electrical Eng., Yonsei Univ.)
  • 발행 : 1991.05.15

초록

In this study electrical impedance plethysmograph was developed and each ten digital blood flow was measured for eight normal males and a patint with digital arterial reconstruction surgery. Blood flow of the digit with surgery was approximately fourty percent compared to that of the other normal digit. For the normal subjects, right-handed subjects generally have larger blood flow in right fingers than that in left fingers while it was reverse for the left-handed subjects. The relative average magnitude of the digital blood flow is the third, first, second, fourth, and fifth digit in descending order for the left and right hand, respectively. Blood flow of the fifth digit for all subjects except one was the smallest among the five digits.