A new in vivo high resolution imaging method which is performed with a newly developed three channel surface gradient coil (SGC) is described. The surface gradient coil can produce more than an order of magnitude stronger gradient fields with good linearity within a limited imaging region. To increase the signal to noise ratio (SNR), we have developed an RF coil integrated surface gradient coil set. In this paper, the geometrical structures and characteristics of the proposed surface gradient coil are discussed and experimentally obtained high resolution images ($50\;{\mu}m$ to $100\;{\mu}m$) of a water filled phantom and a human volunteer's knee using the new surface RF coil integrated SGC set are presented for the demonstration of the in vivo high resolution imaging capability of the new imaging method.