Reduction of Susceptibility Artifact in MR Gradient Echo Imaging

MR Gradient Echo 영상술에서 자화율에 의한 에러의 감소에 관한 연구

  • Ro, Y.M. (Department of Electrical Science, Korea Advanced Institute of Science) ;
  • Cho, Z.H. (Department of Electrical Science, Korea Advanced Institute of Science)
  • 노용만 (한국과학기술원 전기 및 전자공학과) ;
  • 조장희 (한국과학기술원 전기 및 전자공학과)
  • Published : 1990.11.16

Abstract

A new technique for reducing the susceptibility artifact in gradient echo imaging which uses a tailored RF pulse is proposed. It is applied to the case of imaging where artifacts and distortions arise due to the high local magnetic field inhomogeneity i. e., the susceptibility. The signal loss and void phenomena due to susceptibility in a voxel are studied and a correction method using a tailored RF pulse is proposed. Applications of this method in imaging are given and experimental results obtained using an human volunteer with a 2.0 T KAIS NMR system are presented.

Keywords