혈류의 Flow-Turbulence 측정에 관한 연구

A study on the measurement of Blood flow-turbulence

  • 고연순 (아주 대학교 전자공학과) ;
  • 강충신 (아주 대학교 전자공학과) ;
  • 김영길 (아주 대학교 전자공학과)
  • 발행 : 1988.07.01

초록

The tomographic imaging that employs ultrasonic echos has achieved outstanding advances in recent years, and today, ultrasonic diagnostic equipment has become the tool that is absolutely indispensible for clinical operations. Meanwhile, the feasility of measuring blood flow in the heart and vessels by the use of Doppler effect in ultrasonic waves is a well known fact. With respect to the method of blood flow measurment, there are two kinds which employ continous wave and pulse wave doppler system. In this paper, we describe the measurment of Blood flow-turbulence using general purpose Digital Signal Processing Board which had been implemented for the purpose of real-time spectrum analyser. Blood flow-turbulence means the blood-flow behavior. And it's value proportional to the spectrum variance. Therefore mean frequency of blood signal and variance provide useful diagnostic information. We have applied to the major arteries and vein, obtained the information about the time dependent blood-flow behavior.